Hi GO! I have a question.I to do the calibration, .I measuse the resistance with 100Hz measuring frequency. It is ok maybe, but with 1kHz not ok.much more the deviation. See in the attached picture. |
|
|
san398: See in the attached picture |
|
|
The -5V is ok (-4.98V).I made the Baton version (power supply with 34063, not ICL7660) Any idea? |
|
|
I measured this values: |
|
|
Connect the resistor 1.5MOhm and test the entire circuit using the oscilloscope at frequencies of 100Hz and 1kHz. Carefully check the form and magnitude of sine, noise, offset, etc. Based on the values (1500000 ----- 1520000 ----- 2760000) the difference should be noticeable. |
|
|
I found one error. The C2 (5.1pF) was not good value. I change it. In the C2 position 4,7pF is good value?.The measured values are better, but i think are not good. I check the frequency, 100Hz and 1kHz ara not stable and sinus is noisy, the frequency to change between 92-107Hz and 957-1057Hz.The frequency stability are ok? Other meter 100Hz 1kHz Thanks GO for your help! Any idea! |
|
|
The frequency of sinus determined the frequency of a quartz resonator Gr1 and should be stable! Better to measure the frequency at the output DA1.1 pin 7. When measuring large resistors are included scaling amplifier (DA1.2), check it. And turn off or remove possible sources of interference (the computer keyboard, lamps, etc.) |
|
|
I changed the quartz, results is same.I change the DA1, results is same.I change the DA5, results is same.I switch off everything that is not necessary, only switch on the scope and the power supply. |
|
|
Which is the best variation? I attached the picture. |
|
|
Forums that do not open the picture... |
|
|